- A to D:
- Voltage range:
- Frequency response:
- Resolution:
- Accuracy:
- Sampling rate:
- Anti-alias filter:
- Additional filtering:
|
- 16 bits (1 part in 65,536)
- 10 mV pk-pk maximum
- 0.05 Hz – 150 Hz
- 0.15 uV
- ±5% of measured voltage
- 600 Hz
- 150 Hz 6-Pole Bessel
- 50 Hz notch, 25 Hz & 35 Hz artefact low pass filter.
Baseline wander correction and R-R interval are performed in the digital domain
|
- A to D:
- Voltage range:
- Sensitivity:
- Frequency response:
- Resolution:
- Accuracy:
- Sampling rate:
- Anti-alias filter:
|
- 10 bits (1 part in 1,024)
- 20 mV – 400 mV maximum
- 4.5 mV/mm
- 0.01 Hz – 10 Hz
- 0.4 mV
- ±5% of measured voltage
- 40 Hz
- 10 Hz 6-Pole Bessel (gain at 40 Hz, –45 dB)
|
- A to D:
- Range:
- Span:
- Resolution:
- Accuracy:
- Sampling rate:
|
- 16 bits (1 part in 65,636)
- 10 – 50 °C
- 40 °C
- 0.001 °C
- ±0.05 °C
- 1 s/s maximum
|
- Measurement range:
- Technique:
- Accuracy:
|
- 20 – 260 mmHg
- Oscillometry with step deflation
- Meets ANSI/AAMI SP10-1992 Pulse rate ±2% or ±3
|
- Oxygen saturation range:
- Pulse rate range:
- Measurement wavelengths:
- Using Nonin & Nellcor sensors:
Accuracy:
- SpO2:
|
- 0 to 100%
- 18 to 300 pulses per minute
- Red: 660 nanometers @ 3 mW Nominal
- Infrared: 910 nanometers @ 3 mW Nominal
-
| 70 – 100% |
|
± 1 SD |
| No Motion |
(Adults, Pediatrics) |
± 2 digits |
| Motion |
(Adults, Pediatrics) |
± 2 digits |
| Low Perfusion |
(Adults, Pediatrics) |
± 2 digits |
|